[1]于贝贝 荆林凯 王 劲 王贵怀.神经外科手术机器人置入椎弓根螺钉的精准性[J].中国临床神经外科杂志,2021,26(05):360-362.[doi:10.13798/j.issn.1009-153X.2021.05.013]
 YU Bei-bei,JING Lin-kai,WANG Jin,et al.Accuracy of neurosurgical robot in pedicle screw placement[J].,2021,26(05):360-362.[doi:10.13798/j.issn.1009-153X.2021.05.013]
点击复制

神经外科手术机器人置入椎弓根螺钉的精准性()
分享到:

《中国临床神经外科杂志》[ISSN:1009-153X/CN:42-1603/TN]

卷:
26
期数:
2021年05期
页码:
360-362
栏目:
实验研究
出版日期:
2021-05-25

文章信息/Info

Title:
Accuracy of neurosurgical robot in pedicle screw placement
文章编号:
1009-153X(2021)05-0360-03
作者:
于贝贝 荆林凯 王 劲 王贵怀
100084 北京,清华大学临床医学院(于贝贝);102218 北京,清华大学附属北京清华长庚医院神经外科(荆林凯、王 劲、王贵怀)
Author(s):
YU Bei-bei1 JING Lin-kai2 WANG Jin2 WANG Gui-huai2.
1. School of Clinical Medicine, Tsinghua University, 100084 Beijing, China; 2. Department of Neurosurgery, Changgung Hospital, Clinical Institute for Neuroscience, Medical Center, Tsinghua University,Beijing 102218, China
关键词:
脊柱手术手术机器人椎弓根螺钉螺钉置入术
Keywords:
Spine surgery Surgical robot Pedicle screw placement Precision
分类号:
R 651.1+1
DOI:
10.13798/j.issn.1009-153X.2021.05.013
文献标志码:
A
摘要:
目的 探讨神经外科手术机器人在人体腰椎模型进行椎弓根螺钉置入的精准性。方法 利用O型臂影像系统对模型进行正侧位扫描,并三维重建,获取3D-CT数据,传入手术机器人系统,规划椎弓根螺钉的最佳进钉点和进钉方向。手术机器人系统利用3D-CT数据自动注册后,置入椎弓根螺钉。应用置钉前后的3D融合图像,按照Gertzbein-Robbins分级评估置钉的准确性。结果 共置入39枚椎弓根螺钉,其中37枚螺钉完全位于椎弓根内,2枚螺钉突破椎弓根的内壁2 mm以内。螺钉与原规划钉道的平均偏移距离为(1.45±0.67)mm。螺钉在进钉点的滑移方向:内下方16枚(41.03%),外上方6枚(15.38%),下方6枚(15.38%),外下方5枚(12.82%),内上方4枚(10.26%),内方1枚(2.56%),上方1枚(2.56%)。结论 神经外科手术机器人置入椎弓根螺钉具有较高的精准性。
Abstract:
Objective To explore the accuracy of pedicle screw placement by neurosurgical robot in the human lumbar spine model. Methods The front- side-view scan and three-dimensional reconstruction of the human lumbar spine model was performed on the O-arm image system to obtain 3D-CT data which were input into the surgical robotic system to to plan the best insertion point and direction of pedicle screw placement. The surgical robotic system registered automatically using these 3D-CT data and then placed the pedicle screws. The accuracy of pedicle screw placement was evaluated according to the Gertzbein-Robbins classification by the 3D fused image before and after the placement. Results A number of 39 pedicle screws was placed by the surgical robotic system in this study. Of these 39 pedicle screws, 37 screws were completely located in the pedicle and 2 broke through the inner wall of the pedicle within 2 mm; 16 (41.03%) screws slided inward and downward at the entry point, 6 outward and upward, 6 downward, 5 outward and downward 4 inward and upward, 1 inward and 1 upward. The average deviation distance between the screw and the original planned path was (1.45±0.67) mm. Conclusions The neurosurgical robot has high precision in the pedicle screw placement.

参考文献/References:

[1] Nasser R, Yadla S, Maltenfort MG, et al. Complications in spine surgery [J]. J Neurosurg Spine, 2010, 13(2): 144-157.
[2] Akins PT, Harris J, Alvarez JL, et al. Risk factors associatedwith 30-day readmissions after instrumented spine surgery in 14,939 patients: 30-day readmissions after instrumented spine surgery [J]. Spine (Phila Pa 1976), 2015, 40(13): 1022-1032.
[3] Gertzbein SD, Robbins SE. Accuracy of pedicular screw placement in vivo [J]. Spine (Phila Pa 1976), 1990, 15(1): 11-14.
[4] Fujishiro T, Nakaya Y, Fukumoto S, et al. Accuracy of pedi-cle screw placement with robotic guidance system: a cada-veric study [J]. Spine (Phila Pa 1976), 2015, 40(24): 1882-1889.
[5] Lefranc M, Peltier J. Accuracy of thoracolumbar transpedi-cular and vertebral body percutaneous screw placement: coupling the Rosa(R) Spine robot with intraoperative flat-panel CT guidance--a cadaver study [J]. J Robot Surg, 2015, 9(4): 331-338.
[6] Jiang B, Karim Ahmed A, Zygourakis CC, et al. Pedicle screw accuracy assessment in ExcelsiusGPS? robotic spine surgery: evaluation of deviation from pre-planned trajectory [J]. Chin Neurosurg J, 2018, 4(1):1-6.
[7] Han X, Tian W, Liu Y, et al. Safety and accuracy of robot-assisted versus fluoroscopy-assisted pedicle screw insertionin thoracolumbar spinal surgery: a prospective randomized controlled trial [J]. J Neurosurg Spine, 2019, 30(5): 615-622.
[8] Kim HJ, Jung WI, Chang BS, et al. A prospective, randomi-zed, controlled trial of robot-assisted vs freehand pedicle screw fixation in spine surgery [J]. Int J Med Robot, 2017, 13(3): e1779.
[9] Ringel F, Stuer C, Reinke A, et al. Accuracy of robot-assist-ed placement of lumbar and sacral pedicle screws: a pros-pective randomized comparison to conventional freehand screw implantation [J]. Spine (Phila Pa 1976), 2012, 37(8): e496-501.
[10] Urakov TM, Chang KH, Burks SS, et al. Initial academic experience and learning curve with robotic spine instru-mentation [J]. Neurosurg Focus, 2017, 42(5): e4.

相似文献/References:

[1]张圣坤 综述 初 明 审校.重建椎管稳定性的颈椎椎板成形术的研究进展[J].中国临床神经外科杂志,2020,(07):488.[doi:10.13798/j.issn.1009-153X.2020.07.026]

备注/Memo

备注/Memo:
通讯作者:王贵怀,E-mail:guihuai_wang@163.com
更新日期/Last Update: 2021-05-25